How
to maintain a dental handpiece
It is better to alternate operate two high speed handpiece,each
dental unit should have two pcs high speed handpiece and one piece low speed handpiece,you’d better
to use one for 10 mins and change to another one,that will be less the “tired”of
the bearing,and longer the life span of the
bearing.it is said 60% bearing of high speed handpiece was damaged by
the “tired”
Clearing and oiling is the most important parts of the handpiece maintain.oiling
is in order to less the friction of ball and bearing so that the handpiece can
rotate in high speed.we need to use the professional handpiece oil to oiling.it
was run away during the operation,so it need to oiling every day to keep the
longer life span
Clean the external side and led tubes using 70% GL alcohol or other
disinfecting agent without having an ammonium or acid basis or chlorine.
Do never immerge the handpiece in disinfection baths.
Lubricate the turbine before each sterilization cycle.Handpieces which are
not subject to a sterilization process must be lubricated before initiating
each work period(in the moring,at moon,and evening).
Sterilize the handpiece and bur changer in autoclave up to 135
centigarde.20 mins with 121 centigrade and 15 mins with 135 centigrade;withdraw
handpiece from autoclave after the sterilization.
Adjusted the different air pressure according to the different brand and
type handpiece.
It need to be professional maintenance when the handpiece was operated
about 3 months.the maintenance should be operated by some professional staff
according to the certain steps.